Jump to content

Confocal endoscopy

From Wikipedia, the free encyclopedia
Confocal endoscopy
Insertion tip of an endoscope

Confocal endoscopy, or confocal laser endomicroscopy (CLE), is a modern imaging technique that allows the examination of real-time microscopic and histological features inside the body. In the word "endomicroscopy", endo- means "within" and -skopein means "to view or observe". CLE, also known as "optical biopsy", can analyse histology and cytology features of a tissue which otherwise is only possible by tissue biopsy. Similar to confocal microscopy, the laser in CLE filtered by the pinhole excites the fluorescent dye through a beam splitter and objective lens. The fluorescent emission then follows similar paths into the detector. A pinhole is used to select emissions from the desired focal plane. Two categories of CLE exist, namely probe-based (pCLE) and the less common endoscopy-based endoscopy (eCLE).[1]

CLE can be intubated to study the gastrointestinal (GI) tract and accessory digestive organs with a fluorescent dye. A variety of diseases, including inflammatory bowel disease (IBD) and Barrett's oesophagus, can be diagnosed by the magnified and in-depth view in combination with traditional endoscopy.[2]

Significance

[edit]

CLE can identify the lesions with a small depths of view under the tissue, in contrast to the surface level in conventional endoscopy.[3] It also allows clinicians to discriminate benign or malignant lesions through real-time histological diagnosis by revealing the properties of the lamina at a cellular level.[3][4]

An example is Whipple's disease.[5] Conventional endoscopy presents a whitish-patterned duodenal mucosa. CLE, in comparison, generates two images –– the superficial images show capillary leak in duodenal mucosa while the deep images show cells of duodenal mucosa, including goblet cells and foamy macrophages in lamina propria. Compared to histological examination of the same duodenal site after periodic acid-Schiff staining, CLE identifies similar patterns of goblet cells and foamy macrophages.[5]

Types

[edit]

Two types of CLE have been invented, namely probe-based (pCLE) and endoscope-based CLE (eCLE).

Probe-based CLE

[edit]

pCLE, developed by Mauna Kea Technologies, is a fibre bundle transit through the 2.8 mm working channel (the hollow hole) of the standard endoscope into the GI tract.[1] With a fixed plane of imaging, each fibre acts as a pinhole to filter unwanted noise. The frame rate lies between 9 and 12 images/second.[1]

pCLE Products Usage
GastroFlex Upper GI
ConolFlex Lower GI
CholangioFlex Cholangiopancreatography
AQ-19 Fine needle aspiration
Needle-based CLE Endoscopic ultrasound needle (EUS)

Endoscope-based CLE

[edit]

eCLE, developed by Pentax, is a confocal microscopy fixed at the end of the endoscopic tube. The integrated machine of eCLE is larger than the pCLE in diameter, making GI tract endoscopic intubation more difficult. eCLE has ceased commercially due to the camera's inflexibility.[1]

Medical uses

[edit]
Colorectal cancer

Oesophagus

[edit]

CLE is effective in detecting premalignant, including Barrett's oesophagus, and malignant (cancerous) lesions in the upper GI tract.[3][6][7] The modifications of mucosa shown in histopathology as an index of malignancy can be identified under CLE, such as high-grade dysplasia. CLE can also be implemented to refer to the treatment of Barrett's oesophagus by measuring the lateral extent of neoplasia.[3][7]

The Miami classification is the most popular system in oesophageal CLE diagnosis.[6][8]

Stomach and duodenum

[edit]

Similar to that of the oesophagus, CLE is able to detect early gastric cancer, as well as premalignant conditions, such as gastritis and intestinal metaplasia.[1][2][6] CLE can detect and distinguish the stomach pit patterns to identify the disease in accordance with the Miami classification, which was refined in 2016 to include both pit patterns and the architecture of blood vessels.[1][2][8][9] The refined classification allows clinicians to differentiate between neoplastic and non-neoplastic lesions.[9]

The presence of Helicobacter pylori can also be identified using CLE by viewing the morphological changes in tissues.[10][11]

Lower Gastrointestinal Tract

[edit]

CLE reveals "soccer ball-like pattern" of narrower capillaries in malignant lymphomas; distorted architecture and fluorescein leakage from lumen in colonic adenocarcinoma; and blunt-shaped villi and crypts and increased intraepithelial lymphocytes in coeliac disease.[1][2][12]

CLE can be utilized to identify adenoma and neoplasia in colorectal polyps and lesions.[13][14] The Miami classification provides guidelines for clinicians to differentiate neoplastic and non-neoplastic lesions.[8]

Inflammatory Bowel Disease (IBD)

[edit]

CLE can be used for the identification of IBD and its subtypes (Crohn's disease and ulcerative colitis) based on the observation on morphological characteristics, such as architectural distortion, lowered crypt density, crypt irregularity and an abnormally high density of epithelial gaps.[1][2][15] The prediction of IBD progression on non-inflamed epithelium is achievable, too, making way for a novel "treat-to-target" therapeutic approach.[1][2]

Pancreas

[edit]

Incorporating an EUS, CLE can accurately diagnose pancreatic cystic lesions, including mucinous and non-mucinous lesions.[2][16] Special needles are used to collect fluid and cyst wall tissues for testing.[16] Pancreatic ductal adenocarcinoma (PDAC) can also be viewed by CLE.[4] Observing cystic lesions and PDAC, clinicians can identify early chronic pancreatitis and determine the malignancies of lesions.[4]

Biliary duct

[edit]

Biliary stricture can be viewed by CLE.[1] The Miami and Paris classifications can be adapted to differentiate cancerous and inflammatory causes.[17]

Others

[edit]

The discrimination of inflammation and malignant tumor in lung and the urinary system may be done by using CLE and this is currently under research.[10][18] Some usages such as oral and other head-and-neck cancer diagnosis have been proposed.[19][20]

Molecular imaging

[edit]
Tumour angiogenesis (formation of blood vessels) by VEGF

Antibodies of molecular targets are used to diagnose GI diseases by histology.[21][22] CLE captures the fluorescence produced by specific antibodies binding to vascular endothelial growth factor (VEGF). Comparing the significant difference in fluorescent strength, clinicians can differentiate normal and neoplastic tissue. Molecular imaging with antibodies may be applied to CLE as a diagnostic benchmark due to high correlation with ex vivo microscopy.[21]

The molecular imaging technique can be used in a similar manner in the examination of head and neck cancer using CLE, though the diagnostic targets may be different from those in the gastrointestinal tract.[20][23]

Mechanism

[edit]
Principle of a confocal point sensor –– 1) light source, 2) beam splitter, 3) specimen, 4) focal plane, 5) receiver pinhole, 6) photo detector, 7) illumination pinhole, 8) illumination beam, 9) observation beam from focal plane, 10) observation beam out of focal plane

Basic mechanism

[edit]

The laser emitted by CLE through a pinhole is reflected by the beam splitter or a dichroic mirror and focused by an objective lens. The fluorescent dye in targeted tissue is excited and emits a specific wavelength. The emission from the focal plane of the tissue then is collected by the objective lens and the beam splitter. The laser is eventually filtered by a pinhole to reduce out-of-focus noise to enter the detector or photomultiplier tube.[1]

Topical dyes

[edit]

Cresyl violet and acriflavine can be used as topical dyes. Cresyl violet is a common stain in histology used for light microscopy sections, especially brain sections. In CLE, it can enhance the viewing of the cytoplasm, yet it limits tissue penetration and does not show anything about vasculature. Acriflavine is an antiseptic and dye. In CLE, it can stain the nuclei of GI surface epithelial cells. It is however subjected to cytotoxic and mutagenic properties, in addition to common side effects of irritation.[1]

Intravenous dyes

[edit]
Chemical structure of fluorescein

Fluorescein is the most popular IV dye for CLE. Fluorescein is an FDA-cleared dye that is used in ophthalmology clinics in routine as it appears green under cobalt blue light.[12] It is commonly applied topically to identify corneal diseases with slit lamp microscopes including corneal abrasion, ulcers, and infections; or intravenously to identify retinal diseases with angiography including macular degeneration and diabetic retinopathy. In CLE, it is usually administered intravenously immediately before the intubation of an endoscopic tube. The fluorescence is reported to be the most prominent from a few seconds to 8 minutes.[1] Fluorescein is slowly eliminated; thus the fluorescence slowly decays up to a minimal detectable level after 1 hour, giving a time window for clinicians to investigate.[1]

Recognition and optical flow algorithm

[edit]

CLE's narrow field of vision makes it difficult for clinicians to identify the location and path of the probe, making it challenging to correspond the image obtained and the lesion location and direction. Research has proposed a crypt recognition algorithm, which predicts the pixel displacement by the moving angle and distance. By restoring the exploration path of CLE, clinicians can locate the sites of interest and improve diagnostic efficiency.[24]

Image quality assessment

[edit]

Research has proposed a new assessment method for filtering images yielded from CLE. As CLE often encounters image distortions, the degradation of image quality and loss of image information, eventually increase the difficulty of accurate diagnosis.[25] A new image quality assessment (IQA) utilising Weber's Law and local descriptors assesses the quality and filters images with low diagnostic value.[25]

Limitations

[edit]
Endoscopy with biopsy forceps in the working channel (the hollow tube)

The variety of pathology conditions identifiable by CLE is limited.[3] The histological diagnosis is limited to cancerous lesions and inflammation where the number of specific diseases identifiable is not numerous. Moreover, it requires specific training to operate CLE and correctly interpret CLE images, which are rarely used skills by experts in endoscopy.[10][14] Owing to the narrow field of view, the applications of CLE might be restricted.[1] Computer-aided diagnosis with AI technology may be beneficial in diagnosing CLE images.[2]

Fluorescein is the only safe dye approved while cresyl violet and acriflavine are commonly used agents. The lack of choice of contrast agents may limit the application of CLE.[2] For instance, patients allergic to fluorescein should never undergo CLE that involves the use of this intravenous dye.[4]

The optical system consists of complex microscopic optical instruments, which are difficult to manufacture and assemble.[2] Therefore, the tool is expensive.[3]

CLE is mostly used in combination with other techniques instead of replacing conventional endoscopy with biopsy.[7] CLE can only serve as a complementary to the traditional biopsy. By sharing the same working channel, conventional biopsy and CLE can be done alternatively by single intubation.[2]

Adverse effects

[edit]
Intubation of endoscope

The allergic properties of fluorescein, the common intravenous fluorescent dye for CLE, is the major culprit for the mild adverse events.[13]

Intubation

[edit]

CLE, similar to other diagnostic endoscopic techniques, may give rise to pancreatitis when used to examine the pancreas.[13][26] The likelihood for pancreatitis is especially high in needle-based CLE. The incidence can be minimized by shortening the inspection time and avoiding excessive needle movement within the pancreatic cyst wall.[13]

Fluorescein

[edit]
Skin prick testing for allergies

Mild side effects, which are rare, include[1]

These effects are manageable unless patients have prior experiences of them.[1]

Cases of anaphylaxis are reported by ophthalmological uses of fluorescein. Prophylactic use of antihistamines can reduce the chances of allergic reactions or skin prick tests can identify the risk of allergic reactions.[1]

Acriflavine

[edit]

Acriflavine, another contrast agent for CLE, is potentially carcinogenic to humans due to its known mutagenic ability. The dye is therefore not approved by the FDA.[13]

History

[edit]

CLE is a modern, in vivo adaptation of confocal microscopy, the microscopic technique invented by Marvin Minsky in 1957.[10]

Since 2004, CLE has been used for observing histopathological changes in gastrointestinal tissues.[10]

See also

[edit]

References

[edit]
  1. ^ a b c d e f g h i j k l m n o p q r Pilonis, Nastazja Dagny; Januszewicz, Wladyslaw; di Pietro, Massimiliano (2022). "Confocal laser endomicroscopy in gastro-intestinal endoscopy: technical aspects and clinical applications". Translational Gastroenterology and Hepatology. 7: 7. doi:10.21037/tgh.2020.04.02. PMC 8826043. PMID 35243116.
  2. ^ a b c d e f g h i j k Zhang, Huahui; He, Zhongyu; Jin, Ziyi; Yan, Qinglai; Wang, Peng; Ye, Xuesong (2022). "The development and clinical application of microscopic endoscopy for in vivo optical biopsies: Endocytoscopy and confocal laser endomicroscopy". Photodiagnosis and Photodynamic Therapy. 38: 102826. doi:10.1016/j.pdpdt.2022.102826. PMID 35337998.
  3. ^ a b c d e f Fugazza, Alessandro; Gaiani, Federica; Carra, Maria Clotilde; Brunetti, Francesco; Lévy, Michaël; Sobhani, Iradj; Azoulay, Daniel; Catena, Fausto; de'Angelis, Gian Luigi; de'Angelis, Nicola (2016). "Confocal Laser Endomicroscopy in Gastrointestinal and Pancreatobiliary Diseases: A Systematic Review and Meta-Analysis". BioMed Research International. 2016: 1–31. doi:10.1155/2016/4638683. ISSN 2314-6133. PMC 4773527. PMID 26989684.
  4. ^ a b c d Ge, Qi-Chao; Dietrich, Christoph F; Bhutani, Manoop S; Zhang, Bao-Zhen; Zhang, Yue; Wang, Yi-Dan; Zhang, Jing-Jing; Wu, Yu-Fan; Sun, Si-Yu; Guo, Jin-Tao (2021). "Comprehensive review of diagnostic modalities for early chronic pancreatitis". World Journal of Gastroenterology. 27 (27): 4342–4357. doi:10.3748/wjg.v27.i27.4342. ISSN 1007-9327. PMC 8316907. PMID 34366608.
  5. ^ a b Dolak, W.; Leitner, J.; Maresch, J.; Wrba, F.; Mueller, C. (2010). "In vivo identification by confocal laser endoscopy of foamy macrophages associated with Whipple's disease". Endoscopy. 42 (S 02): E310–E311. doi:10.1055/s-0030-1255809. ISSN 0013-726X. PMID 21113886.
  6. ^ a b c Han, Wei; Kong, Rui; Wang, Nan; Bao, Wen; Mao, Xinli; Lu, Jie (2023). "Confocal Laser Endomicroscopy for Detection of Early Upper Gastrointestinal Cancer". Cancers. 15 (3): 776. doi:10.3390/cancers15030776. ISSN 2072-6694. PMC 9913498. PMID 36765734.
  7. ^ a b c Vranić, Luka; Nadarević, Tin; Štimac, Davor (2022). "Probe-Based Confocal Laser Endomicroscopy and Barrett's Esophagus: Just a Scientific Toy or Significant Improvement in Diagnosis?". Digestive Diseases. 40 (1): 97–105. doi:10.1159/000516257. ISSN 0257-2753. PMID 33794523.
  8. ^ a b c Wallace, M.; Lauwers, G.; Chen, Y.; Dekker, E.; Fockens, P.; Sharma, P.; Meining, A. (2011). "Miami classification for probe-based confocal laser endomicroscopy". Endoscopy. 43 (10): 882–891. doi:10.1055/s-0030-1256632. ISSN 0013-726X. PMID 21818734.
  9. ^ a b Canakis, Andrew (2022). "The diagnostic performance of probe-based confocal laser endomicroscopy in the detection of gastric cancer: a systematic review and meta-analysis" (PDF). Annals of Gastroenterology. 35 (5): 496–502. doi:10.20524/aog.2022.0741. PMC 9399575. PMID 36061161.
  10. ^ a b c d e Balañà, Josep; Territo, Angelo; Gallioli, Andrea; Sanguedolce, Francesco; Subiela, José Daniel; Mercadé, Asier; Girón, Irene; Mosquera, Lucía; Gaya, J. M.; Palou, Joan; Breda, Alberto (2019). "Confocal laser endomicroscopy for the treatment of upper urinary tract transitional cellcarcinoma". Archivos Espanoles de Urologia (in Spanish). 72 (8): 765–771. ISSN 0004-0614. PMID 31579035.
  11. ^ Yang, Hang; Hu, Bing (2021). "Diagnosis of Helicobacter pylori Infection and Recent Advances". Diagnostics. 11 (8): 1305. doi:10.3390/diagnostics11081305. ISSN 2075-4418. PMC 8391489. PMID 34441240.
  12. ^ a b Săftoiu, Adrian, ed. (2023). Pocket Guide to Advanced Endoscopy in Gastroenterology. Cham: Springer International Publishing. doi:10.1007/978-3-031-42076-4. ISBN 978-3-031-42075-7.
  13. ^ a b c d e Al-Mansour, Mazen R.; Caycedo-Marulanda, Antonio; Davis, Brian R.; Alawashez, Abdulrahim; Docimo, Salvatore; Qureshi, Alia; Tsuda, Shawn (2021). "SAGES TAVAC safety and efficacy analysis confocal laser endomicroscopy". Surgical Endoscopy. 35 (5): 2091–2103. doi:10.1007/s00464-020-07607-3. ISSN 0930-2794. PMID 32405892.
  14. ^ a b Glover, Ben; Teare, Julian; Patel, Nisha (2020). "The Status of Advanced Imaging Techniques for Optical Biopsy of Colonic Polyps". Clinical and Translational Gastroenterology. 11 (3): e00130. doi:10.14309/ctg.0000000000000130. ISSN 2155-384X. PMC 7145035. PMID 32352708.
  15. ^ Buchner, Anna M (2019). "Confocal Laser Endomicroscopy in the Evaluation of Inflammatory Bowel Disease". Inflammatory Bowel Diseases. 25 (8): 1302–1312. doi:10.1093/ibd/izz021. ISSN 1078-0998. PMID 30877772.
  16. ^ a b Kim, Do Han; Krishna, Somashekar G.; Coronel, Emmanuel; Kröner, Paul T.; Wolfsen, Herbert C.; Wallace, Michael B.; Corral, Juan E. (2022). "Confocal Laser Endomicroscopy in the Diagnosis of Biliary and Pancreatic Disorders: A Systematic Analysis". Clinical Endoscopy. 55 (2): 197–207. doi:10.5946/ce.2021.079. ISSN 2234-2400. PMC 8995979. PMID 34839621.
  17. ^ Zhou, Selena; Buxbaum, James (2022). "Advanced Imaging of the Biliary System and Pancreas". Digestive Diseases and Sciences. 67 (5): 1599–1612. doi:10.1007/s10620-022-07389-9. ISSN 0163-2116. PMID 35171407.
  18. ^ Xia, Yang; Li, Qin; Zhong, Changgao; Wang, Kopen; Li, Shiyue (2023). "Inheritance and innovation of the diagnosis of peripheral pulmonary lesions". Therapeutic Advances in Chronic Disease. 14: 204062232211467. doi:10.1177/20406223221146723. ISSN 2040-6223. PMC 9896091. PMID 36743297.
  19. ^ Farahati, Behnaz; Stachs, Oliver; Prall, Friedrich; Stave, Joachim; Guthoff, Rudolf; Pau, Hans Wilhelm; Just, Tino (2010). "Rigid confocal endoscopy for in vivo imaging of experimental oral squamous intra-epithelial lesions". Journal of Oral Pathology & Medicine. 39 (4): 318–327. doi:10.1111/j.1600-0714.2009.00841.x. ISSN 0904-2512. PMID 20050982.
  20. ^ a b Abbaci, Muriel; Breuskin, Ingrid; Casiraghi, Odile; De Leeuw, Frederic; Ferchiou, Malek; Temam, Stephane; Laplace-Builhé, Corinne (2014). "Confocal laser endomicroscopy for non-invasive head and neck cancer imaging: A comprehensive review". Oral Oncology. 50 (8): 711–716. doi:10.1016/j.oraloncology.2014.05.002. PMID 24932530.
  21. ^ a b Foersch, S.; Kiesslich, R.; Waldner, M. J.; Delaney, P.; Galle, P. R.; Neurath, M. F.; Goetz, M. (2010). "Molecular imaging of VEGF in gastrointestinal cancer in vivo using confocal laser endomicroscopy". Gut. 59 (8): 1046–1055. doi:10.1136/gut.2009.202986. ISSN 0017-5749. PMID 20639250.
  22. ^ Choi, Kwi-Sook; Jung, Hwoon-Yong (2014). "Confocal Laser Endomicroscopy and Molecular Imaging in Barrett Esophagus and Stomach". Clinical Endoscopy. 47 (1): 23–30. doi:10.5946/ce.2014.47.1.23. ISSN 2234-2400. PMC 3928487. PMID 24570880.
  23. ^ Kakaletri, Irini; Linxweiler, Maximilian; Ajlouni, Serine; Charalampaki, Patra (2022). "Development, Implementation and Application of Confocal Laser Endomicroscopy in Brain, Head and Neck Surgery—A Review". Diagnostics. 12 (11): 2697. doi:10.3390/diagnostics12112697. ISSN 2075-4418. PMC 9689276. PMID 36359540.
  24. ^ Yu, Haoxiang; Lu, Yuhua; Liu, Qian (2023). "Ranging of Confocal Endoscopy Probe Using Recognition and Optical Flow Algorithm". In Yu, Chen; Zhou, Jiehan; Song, Xianhua; Lu, Zeguang (eds.). Green, Pervasive, and Cloud Computing. Lecture Notes in Computer Science. Vol. 13744. Cham: Springer International Publishing. pp. 122–131. doi:10.1007/978-3-031-26118-3_9. ISBN 978-3-031-26118-3.
  25. ^ a b Dong, Xiangjiang; Fu, Ling; Liu, Qian (2022-05-18). "No-reference image quality assessment for confocal endoscopy images with perceptual local descriptor". Journal of Biomedical Optics. 27 (5). Bibcode:2022JBO....27e6503D. doi:10.1117/1.JBO.27.5.056503. ISSN 1083-3668. PMC 9116465. PMID 35585672.
  26. ^ Adler, Douglas G. (2015). "Consent, Common Adverse Events, and Post–Adverse Event Actions in Endoscopy". Gastrointestinal Endoscopy Clinics of North America. 25 (1): 1–8. doi:10.1016/j.giec.2014.09.001. PMID 25442954.